Design and implementation of a differential scanning calorimeter for the simultaneous measurement of small angle neutron scattering

No Thumbnail Available
Date
2014-04-08
Journal Title
Journal ISSN
Volume Title
Publisher
IOP Science
Abstract
For almost 30 years, at synchrotron facilities, it has been possible to perform small-angle x-ray scattering experiments whilst simultaneously measuring differential scanning calorimetry (DSC). However, a range of challenges exist to enable the collection of simultaneous small-angle neutron scattering (SANS) and DSC data associated not only with intrinsic flux limitations but also scattering geometry and thermal control. The development of a DSC (temperature range ca. −150 ◦C to 500 ◦C) suitable for SANS is detailed here and its successful use is illustrated with combined measurements on a binary blend of normal alkanes in which one component has been deuterium labelled. © 2014, IOP Publishing Ltd
Description
Keywords
Neutrons, Calorimetry, X-ray diffraction, Data, Deuterium, Temperature range
Citation
Pullen, S. A., Booth, N., Olsen, S. R., Day, B., Franceschini, F., Mannicke, F. & Gilbert, E. P. (2014). Design and implementation of a differential scanning calorimeter for the simultaneous measurement of small angle neutron scattering. Measurement Science and Technology, 25(5), 055606. doi:10.1088/0957-0233/25/5/055606