Evaluation of lead shielding for a gamma-spectroscopy system

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier

Abstract

Three types of lead shielding assemblies were investigated to establish which provides the lowest background as preliminary work for determining the efficiency of a High Purity Germanium (HPGe) detector. The background continuum and peaks were examined using gamma spectroscopy with a coaxial and a planar HPGe detector. The Pb-210 concentration in each lead sample taken from the assemblies was quantitatively determined. A French lead sample was found to have the lowest background in this relative comparison with a Pb-210 concentration of 29.1 +/- 1.3 Bq kg(-1). © 2008, Elsevier Ltd.

Description

Citation

Smith, M. L., Bignell, L., Alexiev, D., Mo, L., & Harrison, J. (2008). Evaluation of lead shielding for a gamma-spectroscopy system. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 589(2), 275-279. doi:10.1016/j.nima.2008.02.050