Influence of rejection of a fraction of the single photoelectron peak in liquid scintillation counting
No Thumbnail Available
Date
2006-03-15
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
Abstract
Zero discrimination level setting of liquid scintillation systems for counting of all single photoelectron pulses is required by both triple-to-double coincidence ratio (TDCR) and CIEMAT/NIST methods. However, this requirement may not be able to be met in some circumstances. The influence of rejection of a fraction of the single photoelectron peak has been studied both theoretically and experimentally on 3H, 63Ni, 14C and 90Sr/90Y for both the TDCR and CIEMAT/NIST methods. A modified formula for calculation of efficiency is given in consideration of the rejected fraction of the single photoelectron peak. © 2006, Elsevier Ltd.
Description
Keywords
Photoelectron counting, Fractionation, Liquid scintillation detectors, Coincidence methods, Peaks, Pulses
Citation
Mo, L., Cassette, P., & Baldock, C. (2006). Influence of rejection of a fraction of the single photoelectron peak in liquid scintillation counting. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 558(2), 490-496. doi:10.1016/j.nima.2005.11.210