Correction for intrinsic and set dead-time losses in radioactivity counting

dc.contributor.authorWyllie, HAen_AU
dc.date.accessioned2007-11-22T04:21:39Zen_AU
dc.date.accessioned2010-04-30T04:30:27Zen_AU
dc.date.available2007-11-22T04:21:39Zen_AU
dc.date.available2010-04-30T04:30:27Zen_AU
dc.date.issued1992-12en_AU
dc.description.abstractEquations are derived for the determination of the intrinsic dead time of the components which precede the paralysis unit in a counting system for measuring radioactivity. The determination depends on the extension of the set dead time by the intrinsic dead time. Improved formulae are given for the dead-time correction of the count rate of a radioactive source in a single-channel system. A variable in the formulae is the intrinsic dead time which is determined concurrently with the counting of the source. The only extra equipment required in a conventional system is a scaler.en_AU
dc.identifier.citationWyllie, H. A. (1992). Correction for intrinsic and set dead-time losses in radioactivity counting (ANSTO/E701). Lucas Heights, NSW:Australian Nuclear Science and Technology Organisation.en_AU
dc.identifier.govdoc221en_AU
dc.identifier.isbn0642599165en_AU
dc.identifier.issn10307745en_AU
dc.identifier.otherANSTO-E-701en_AU
dc.identifier.placeofpublicationLucas Heights, New South Walesen_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/399en_AU
dc.language.isoen_auen_AU
dc.publisherAustralian Nuclear Science and Technology Organisationen_AU
dc.subjectCounting techniquesen_AU
dc.subjectScalersen_AU
dc.subjectMeasuring instrumentsen_AU
dc.subjectDead timeen_AU
dc.titleCorrection for intrinsic and set dead-time losses in radioactivity countingen_AU