Deep level transient conductance spectrometer for high resistivity semiconductors using a marginal oscillator detector.

Loading...
Thumbnail Image
Date
1984-09
Journal Title
Journal ISSN
Volume Title
Publisher
Australian Atomic Energy Commission
Abstract
A deep level transient conductance spectrometer for high resistivity semiconductors, using a radiofrequency (~40 MHz) marginal oscillator as a conductance detector, is described. Spectra are generated by periodically filling deep level trapping centres with carriers stimulated by a pulsed GaAs laser, and processing the trap-emptying conductance signal through an exponential Miller correlator as the sample temperature is slowly ramped. Simple capacitive coupling of samples to the oscillator tank circuit eliminates problems such as unwanted defect annealing and other material changes often associated with the high temperature techniques necessary for ohmic contact formation. Representative deep level spectra are given for semi-insulating Bridgman-grown CdTe.
Description
Keywords
Cadmium tellurides, Oscillators, Semiconductor materials, Spectrometers
Citation
Alexiev, D., Tavendale, A. 1984). A deep level transient conductance spectrometer for high resistivity semiconductors using a marginal oscillator detector (AAEC-E-598). Lucas Heights N.S.W.: Australian Atomic Energy Commission, Research Establishment.