An equation is derived for calculating the radioactivity of a source from the results of coincidence counting taking into account dead-time losses and accidental coincidences. The corrections allow for the extension of the set dead time in the β channel by the intrinsic dead time. Experimental verification shows improvement over a previous equation.
Description
Citation
Wyllie, H. A. (1987). Coincidence-counting corrections for accidental coincidences, set dead time and intrinsic dead time (ANSTO-E-700). Lucas Heights, NSW: Australian Nuclear Science and Technology Organisation.