Correction for the effect of radionuclidic impurities in ionisation chamber measurements of 201Tl

dc.contributor.authorBuckman, SMen_AU
dc.date.accessioned2022-08-10T02:56:26Zen_AU
dc.date.available2022-08-10T02:56:26Zen_AU
dc.date.issued1994-11-16en_AU
dc.date.statistics2022-04-07en_AU
dc.descriptionPhysical copy held by ANSTO Library at DDC 541.382/16.en_AU
dc.description.abstractThe National Medical Cyclotron (NMC) in Sydney commenced commercial production of 201T1 in May I993(l). From the locally produced 202T1, Australian Radioisotopes (ARI) produces radiopharmaceuticals which are supplied to the nuclear medicine community. Under the National Measurement Act, 1960(2) the activities quoted on products supplied by ARI must be legally traceable to National standards of measurement. The Australian primary standard of measurement for radioactivity is the legal responsibility of the Radiation Standards Project at ANSTO(5). The Radiation Standards Project, in collaboration with the Quality Control section at ARI, has established an on-going calibration program to ensure that ARI's activity measuring equipment is traceable to National standards. Soon after the production of 201T1 commenced at the NMC, inconsistencies were observed in ionisation chamber measurements of 201T1 solutions. The cause for these inconsistencies was shown to be the presence of the radionuclidic impurities, 200T1 and 202T1. Even though the levels of these impurities were well below the recommended limits(4), their effect on ionisation chamber measurements could be relatively large. In order to correct for the effect of these impurities, calibration factors for 200T1 and 202T1 were determined using two independent methods. In the first method(5), the calibration factors were calculated from the decay scheme of the nuclides and the response curve of the ionisation chamber. However, this method can only be applied when the response curve of the ionisation chamber is reasonably well known and where there is no significant contribution due to bremsstrahlung. An alternative and more reliable method was developed which used statistical modelling techniques(6) to obtain maximum likelihood estimates of the calibration factors. This method can be used in cases where there may be significant bremsstrahlung production or where the gamma-energy response curve is unknown. By using these methods, calibration factors were determined for two TPA chambers and a Vinten IV dose calibrator. The results from the two methods were in agreement to within the measurement uncertainties. By incorporating these calibration factors into the measurement systems of both the Quality Control and Production sections at ARI, the effect of these impurities on ionisation chamber measurements was successfully eliminated. The accuracy of the Australian standard of measurement for 201T1 has now been confirmed by International intercomparison through the international reference system for gamma-emitting nuclides(7).en_AU
dc.identifier.citationBuckman, S. M. (1994). Correction for the effect of radionuclidic impurities in ionisation chamber measurements of 201Tl. Poster presented to the AINSE Conference on Radiation Biology and Chemistry, Cuming Theatre, Chemistry Department, The University of Melbourne, Melbourne, Victoria, 16-18 November 1994, (pp. 29-30).en_AU
dc.identifier.conferenceenddate18 November 1994en_AU
dc.identifier.conferencenameAINSE Conference on Radiation Biology and Chemistryen_AU
dc.identifier.conferenceplaceMelbourne, Victoriaen_AU
dc.identifier.conferencestartdate16 November 1994en_AU
dc.identifier.isbn9780959847260en_AU
dc.identifier.otherPaper no.14en_AU
dc.identifier.pagination29-30en_AU
dc.identifier.urihttps://apo.ansto.gov.au/dspace/handle/10238/13498en_AU
dc.language.isoenen_AU
dc.publisherAustralian Institute of Nuclear Science and Engineeringen_AU
dc.subjectCalibration standardsen_AU
dc.subjectDataen_AU
dc.subjectIonization chambersen_AU
dc.subjectRadiation detectorsen_AU
dc.subjectRadioisotopesen_AU
dc.subjectThalliumen_AU
dc.subjectCyclotronsen_AU
dc.subjectImpuritiesen_AU
dc.titleCorrection for the effect of radionuclidic impurities in ionisation chamber measurements of 201Tlen_AU
dc.typeConference Posteren_AU
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
27007853.pdf
Size:
2.83 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.63 KB
Format:
Item-specific license agreed upon to submission
Description: