Calibration of temperature controllers and sensors

dc.contributor.authorManning, AGen_AU
dc.contributor.authorBooth, Nen_AU
dc.contributor.authorImperia, Pen_AU
dc.date.accessioned2021-10-27T21:24:18Zen_AU
dc.date.available2021-10-27T21:24:18Zen_AU
dc.date.issued2016-11-29en_AU
dc.date.statistics2021-10-13en_AU
dc.description.abstractReliable temperature measurements are a key part of many sample environment setups for neutron scattering experiments. This typically requires the use of both a temperature sensor and temperature controller. It is often desirable to be able to perform a continuous measurement spanning a wide range of temperatures, however the choice of suitable sensors is limited. Rhodium iron (RhFe) resistive thermometers are best suited to the temperature range of our cryofurnaces (1.5 to 800 K), however their calibration is known to drift when exposed to elevated temperatures. Here, we will present procedures to ensure that both the controller and sensor used in such experiments are performing optimally. Calibrated resistors, which can mimic the behavior of a temperature sensor fixed at a particular temperature, can be used to check the calibration of the readings given by temperature controllers such as Lake Shore 336/340 or Oxford Instruments iTC units. If the controllers are known to be operating to the tolerances specified by their manufacturer, then we can have confidence in the measurements that they make. We can then use our newly verified controllers to assess the performance of RhFe sensors over a range of temperatures from 4 to 800 K in a custom-built calibration rig. This allows us to quantify the behavior of the sensors after they are subject to high temperatures, and to account for shifts in their readings. Using this procedure, we can ensure that the sensors remain reliable after prolonged operational use.en_AU
dc.identifier.citationManning, A., Booth, N., & Imperia, P. (2016). Calibration of temperature controllers and sensors. Paper presented at 13th AINSE-ANBUG Neutron Scattering Symposium, Sydney, NSW, Australia, 29-30 November 2016.en_AU
dc.identifier.conferenceenddate30 November 2016en_AU
dc.identifier.conferencename13th AINSE-ANBUG Neutron Scattering Symposiumen_AU
dc.identifier.conferenceplaceSydney, NSW, Australiaen_AU
dc.identifier.conferencestartdate29 November 2016en_AU
dc.identifier.urihttps://apo.ansto.gov.au/dspace/handle/10238/12134en_AU
dc.language.isoenen_AU
dc.publisherAustralian Institute of Nuclear Science and Engineeringen_AU
dc.subjectNeutron diffractionen_AU
dc.subjectRhodiumen_AU
dc.subjectIronen_AU
dc.subjectThermometersen_AU
dc.subjectFurnacesen_AU
dc.subjectToleranceen_AU
dc.subjectTemperature controlen_AU
dc.subjectCalibrationen_AU
dc.titleCalibration of temperature controllers and sensorsen_AU
dc.typeConference Abstracten_AU
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Abstract_Handbook_v2(1).pdf
Size:
1.87 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.63 KB
Format:
Item-specific license agreed upon to submission
Description: