dc.description.abstractThis work has evaluated synthetic gamma-ray spectra created by the RASE sampler using experimental data. The RASE sampler resamples experimental data to create large data libraries which are subsequently available for use in evaluation of radionuclide identification algorithms. A statistical evaluation of the synthetic energy bins has shown the variation to follow a Poisson distribution identical to experimental data. The minimum amount of statistics required in each base spectrum to ensure the subsequent use of the base spectrum in the generation of statistically robust synthetic data was determined. A requirement that the simulated acquisition time of the synthetic spectra was not more than 4% of the acquisition time of the base spectrum was also determined. Further validation of RASE was undertaken using two different radionuclide identification algorithms. © 2013, Elsevier Ltd.en_AU
dc.identifier.citationFlynn, A., Boardman, D., & Reinhard, M. I. (2013). The validation of synthetic spectra used in the performance evaluation of radionuclide identifiers. Applied Radiation and Isotopes, 77, 145-152. doi:10.1016/j.apradiso.2013.03.089en_AU
dc.identifier.govdoc6104en_AU
dc.identifier.issn0969-8043en_AU
dc.identifier.journaltitleApplied Radiation and Isotopesen_AU
dc.identifier.pagination145-152en_AU
dc.identifier.urihttp://dx.doi.org/10.1016/j.apradiso.2013.03.089en_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/6265en_AU
dc.identifier.volume77en_AU
dc.language.isoenen_AU
dc.publisherElsevieren_AU
dc.subjectGamma spectraen_AU
dc.subjectAlgorithmsen_AU
dc.subjectRadioisotopesen_AU
dc.subjectExperimental dataen_AU
dc.subjectIAEAen_AU
dc.subjectRadiation monitorsen_AU
dc.titleThe validation of synthetic spectra used in the performance evaluation of radionuclide identifiersen_AU
dc.typeJournal Articleen_AU
Files
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:
Collections