Safety interlock and vent system to alleviate potentially dangerous ice blockage of top-loading cryostat sample sticks

dc.contributor.authorPangelis, Sen_AU
dc.contributor.authorOlsen, SRen_AU
dc.contributor.authorScherschligt, Jen_AU
dc.contributor.authorLeao, JBen_AU
dc.contributor.authorPullen, SAen_AU
dc.contributor.authorDender, Den_AU
dc.contributor.authorHester, JRen_AU
dc.contributor.authorImperia, Pen_AU
dc.date.accessioned2013-10-14T00:30:30Zen_AU
dc.date.available2013-10-14T00:30:30Zen_AU
dc.date.issued2013-08-01en_AU
dc.date.statistics2013-10-14en_AU
dc.description.abstractA combined solution is presented for minimizing the safety hazards associated with closed cycle cryostats described by Swainson & Cranswick [J. Appl. Cryst. (2010), 43, 206-210]. The initial solution is to install a vent tube with one open end deep inside the sample space and a pressure relief valve at the top. This solution works for either a cryogen or a cryogen-free (closed cycle) system. The second approach, which can be combined with the first and is applicable to cryogen-free cryostats, involves electrically interlocking the closed cycle refrigerator compressor to the sample space, so that the system cannot be cooled in the presence of a leak path to air. © 2013, Wiley-Blackwell.en_AU
dc.identifier.citationPangelis, S., Olsen, S. R., Scherschligt, J., Leao, J. B., Pullen, S. A., Dender, D., Hester, J. R., & Imperia, P. (2013). Safety interlock and vent system to alleviate potentially dangerous ice blockage of top-loading cryostat sample sticks. Journal of Applied Crystallography, 46, 1236-1239. doi:10.1107/S0021889813010868en_AU
dc.identifier.govdoc5071en_AU
dc.identifier.issn0021-8898en_AU
dc.identifier.journaltitleJournal of Applied Crystallographyen_AU
dc.identifier.pagination1236-1239en_AU
dc.identifier.urihttp://dx.doi.org/doi:10.1107/S0021889813010868en_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/4797en_AU
dc.identifier.volume46en_AU
dc.language.isoenen_AU
dc.publisherWiley-Blackwellen_AU
dc.subjectCrystalsen_AU
dc.subjectIceen_AU
dc.subjectFurnacesen_AU
dc.subjectNeutronsen_AU
dc.subjectScatteringen_AU
dc.subjectEnvironmenten_AU
dc.titleSafety interlock and vent system to alleviate potentially dangerous ice blockage of top-loading cryostat sample sticksen_AU