The determination of the efficiency of a Compton suppressed HPGe detector using Monte Carlo simulations

dc.contributor.authorMcNamara, ALen_AU
dc.contributor.authorHeijnis, Hen_AU
dc.contributor.authorFierro, Den_AU
dc.contributor.authorReinhard, MIen_AU
dc.date.accessioned2012-04-18T01:53:02Zen_AU
dc.date.available2012-04-18T01:53:02Zen_AU
dc.date.issued2012-04-01en_AU
dc.date.statistics2012-04-18en_AU
dc.description.abstractCompton suppressed high-purity germanium (HPGe) detector is well suited to the analysis of low levels of radioactivity in environmental samples. The difference in geometry, density and composition of environmental calibration standards (e.g. soil) can contribute to excessive experimental uncertainty to the measured efficiency curve. Furthermore multiple detectors, like those used in a Compton suppressed system, can add complexities to the calibration process. Monte Carlo simulations can be a powerful complement in calibrating these types of detector systems, provided enough physical information on the system is known. A full detector model using the Geant4 simulation toolkit is presented and the system is modelled in both the suppressed and unsuppressed mode of operation. The full energy peak efficiencies of radionuclides from a standard source sample is calculated and compared to experimental measurements. The experimental results agree relatively well with the simulated values (within similar to 5 - 20%). The simulations show that coincidence losses in the Compton suppression system can cause radionuclide specific effects on the detector efficiency, especially in the Compton suppressed mode of the detector. Additionally since low energy photons are more sensitive to small inaccuracies in the computational detector model than high energy photons, large discrepancies may occur at energies lower than similar to 100 keV. Crown Copyright (C) 2011 Published by Elsevier Ltd. All rights reserved.en_AU
dc.identifier.citationMcNamara, A. L., Heijnis, H., Fierro, D., & Reinhard, M. I., (2012). The determination of the efficiency of a Compton suppressed HPGe detector using Monte Carlo simulations. Journal of Environment Radioactivity, 106, 1-7. doi:10.1016/j.jenvrad.2011.10.017en_AU
dc.identifier.govdoc3907en_AU
dc.identifier.issn0265-931Xen_AU
dc.identifier.journaltitleJournal of Environment Radioactivityen_AU
dc.identifier.pagination1-7en_AU
dc.identifier.urihttp://dx.doi.org/10.1016/j.jenvrad.2011.10.017en_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/4147en_AU
dc.identifier.volume106en_AU
dc.language.isoenen_AU
dc.publisherElseiveren_AU
dc.subjectEfficiencyen_AU
dc.subjectRadiation detectorsen_AU
dc.subjectCalibrationen_AU
dc.subjectMonte Carlo Methoden_AU
dc.subjectGamma spectroscopyen_AU
dc.subjectGermaniumen_AU
dc.titleThe determination of the efficiency of a Compton suppressed HPGe detector using Monte Carlo simulationsen_AU
dc.typeJournal Articleen_AU