Repository logo


Comparison of LaBr3 : Ce and LaCl3 : Ce with NaI(Tl) and cadmium zinc telluride (CZT) detectors

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Institute of Electrical and Electronics Engineers (IEEE)

Abstract

Energy resolution and detection efficiency were compared between new scintillators, lanthanum bromide (LaBr3:Ce) and lanthanum chloride (LaCl3:Ce) with conventional detectors, sodium iodide NaI(Tl) and cadmium zinc telluride (CdZnTe or CZT). The study has shown that LaBr3:Ce and LaCl3:Ce scintillator detectors provided by Saint-Gobain offer better resolution than NaI(Tl) detector. LaBr3:Ce and LaCl3:Ce detectors could resolve some closely spaced peaks from 133Ba and 152Eu, which Na(I) could not. LaBr3:Ce has slightly better resolution and higher efficiency than LaCl3:Ce. An overall advantage of LaBr3:Ce and LaCl3:Ce detectors over NaI(Tl) and CZT has been discussed. The intrinsic activity of LaBr3:Ce is also demonstrated in this study. © 2008, Institute of Electrical and Electronics Engineers (IEEE)

Description

Citation

Alexiev, D., Mo, L., Prokopovich, D. A., Smith, M. L., & Matuchova, M. (2008). Comparison of LaBr3 : Ce and LaCl3 : Ce with NaI(Tl) and cadmium zinc telluride (CZT) detectors. IEEE Transactions on Nuclear Science, 55(3), 1174-1177. doi:10.1109/TNS.2008.922837

Collections

Endorsement

Review

Supplemented By

Referenced By