Development of an Australian secondary standard for the reference air kerma rate measurement of 125I seeds

dc.contributor.authorDavies, JBen_AU
dc.contributor.authorMo, Len_AU
dc.contributor.authorAlexiev, Den_AU
dc.date.accessioned2010-08-24T01:13:09Zen_AU
dc.date.available2010-08-24T01:13:09Zen_AU
dc.date.issued2005-09-01en_AU
dc.date.statistics2005-09-01en_AU
dc.description.abstractCalibration of a High-Dose-Rate 1000 Plus ionisation chamber and associated electrometer for the determination of reference air kerma rate for 125I brachytherapy seeds has been developed. Traceability was established from ionisation chamber measurements of an Amersham model 6711125I seed for which a primary standard measurement of reference air kerma rate has been made by the National Institute of Standards and Technology. The status of an Australian secondary standard is to be realised by establishing legal verifying authority from the National Measurement Institute. The calibrated chamber thus provides a measurement from which similar chambers in hospitals may be calibrated. For quality assurance this result was compared with the calibration certificate supplied by the Accredited Dosimetry Calibration Laboratory of the University of Wisconsin, USA. © 2005, Springer.en_AU
dc.identifier.citationDavies, J. B., Mo, L., & Alexiev, D. (2005). Development of an Australian secondary standard for the reference air kerma rate measurement of 125I seeds. Australasian Physical & Engineering Sciences in Medicine, 28(3), 196-199. doi:10.1007/BF03178716en_AU
dc.identifier.govdoc2524en_AU
dc.identifier.issn0158-9938en_AU
dc.identifier.issue3en_AU
dc.identifier.journaltitleAustralasian Physical & Engineering Sciences in Medicineen_AU
dc.identifier.pagination196-199en_AU
dc.identifier.urihttp://dx.doi.org/10.1007/BF03178716en_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/2324en_AU
dc.identifier.volume28en_AU
dc.language.isoenen_AU
dc.publisherSpringeren_AU
dc.subjectBrachytherapyen_AU
dc.subjectIodine 125en_AU
dc.subjectCalibrationen_AU
dc.subjectDosimetryen_AU
dc.subjectIonization chambersen_AU
dc.subjectKermaen_AU
dc.titleDevelopment of an Australian secondary standard for the reference air kerma rate measurement of 125I seedsen_AU
dc.typeJournal Articleen_AU