High sensitivity two filter radon/thoron detectors with a wire or nylon screen as a second filter.

dc.contributor.authorWhittlestone, Sen_AU
dc.contributor.authorZahorowski, Wen_AU
dc.contributor.authorWasiolek, Pen_AU
dc.date.accessioned2007-11-22T04:20:19Zen_AU
dc.date.accessioned2010-04-30T04:30:13Zen_AU
dc.date.available2007-11-22T04:20:19Zen_AU
dc.date.available2010-04-30T04:30:13Zen_AU
dc.date.issued1994-12en_AU
dc.description.abstractA study is made of the use of wire and nylon screens as a second filter in two radon or thoron detectors. It is shown that acceptable detection efficiency is obtained at flow rates comparable to those used in detectors in which other types of filter are used. The main advantage of the screens is their very low flow impedance. Several designs of detector which exploit this feature are discussed. Details are given of the performance of three prototypes: a 32 L radon detector with a limit of detection of 0.0027 Bq m-3 and power consumption of 25 watts; and a portable thoron emanometer capable of detecting fluxes as low as 1 m Bq m -2 s-1. The radon detectors are rugged and simple. They can operate with no routine maintenance and are suited to remote locations where only infrequent technical support is available.en_AU
dc.identifier.citationWhittlestone, S, Zahorowski, W., & Wasiolek, P. (1994). High sensitivity two filter radon/thoron detectors with a wire or nylon screen as a second filter. (ANSTO/E718). Lucas Heights, NSW: Australian Nuclear Science and Technology Organisationen_AU
dc.identifier.govdoc207en_AU
dc.identifier.isbn0642599562en_AU
dc.identifier.issn10307745en_AU
dc.identifier.otherANSTO-E-718en_AU
dc.identifier.placeofpublicationLucas Heights, New South Walesen_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/376en_AU
dc.language.isoen_auen_AU
dc.publisherAustralian Nuclear Science and Technology Organisationen_AU
dc.subjectRadonen_AU
dc.subjectRadon 220en_AU
dc.subjectRadiation detectorsen_AU
dc.subjectFiltersen_AU
dc.titleHigh sensitivity two filter radon/thoron detectors with a wire or nylon screen as a second filter.en_AU