Full-energy-peak efficiencies of three gamma-ray detectors

dc.contributor.authorHanna, GLen_AU
dc.contributor.authorWalker, DGen_AU
dc.contributor.authorBeach, PMen_AU
dc.date.accessioned2007-11-22T04:30:47Zen_AU
dc.date.accessioned2010-04-30T04:40:38Zen_AU
dc.date.available2007-11-22T04:30:47Zen_AU
dc.date.available2010-04-30T04:40:38Zen_AU
dc.date.issued1967-01en_AU
dc.description.abstractTwo sodium iodide scintillators and a lithium-drifted germanium detector are used in the analysis by gamma-ray spectrometry of gaseous fission products obtained in sweep-capsule fission product release experiments. A description is given of the measurements of full-energy-peak efficiencies of the three detectors for the source geometries used in counting the fission product samples. Experimental efficiencies are compared with calculated efficiencies for one of the sodium iodide detectors. Measurements made of the resolutions obtained with each detector are also noted.en_AU
dc.identifier.citationHanna, G. L., Walker, D. G., & Beach, P. M., & Australian Atomic Energy Commission (1967). Full-energy-peak efficiencies of three gamma-ray detectors (AAEC/TM384). Lucas Heights N.S.W. Research Establishment: Australian Atomic Energy Commission.en_AU
dc.identifier.govdoc811en_AU
dc.identifier.otherAAEC-TM-384en_AU
dc.identifier.placeofpublicationLucas Heights, New South Walesen_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/801en_AU
dc.language.isoen_auen_AU
dc.publisherAustralian Atomic Energy Commissionen_AU
dc.subjectSelf-powered gamma detectorsen_AU
dc.subjectGE semiconductor detectorsen_AU
dc.subjectNAI detectorsen_AU
dc.subjectScintillationsen_AU
dc.subjectFission productsen_AU
dc.titleFull-energy-peak efficiencies of three gamma-ray detectorsen_AU
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
AAEC-TM-384.pdf
Size:
435.27 KB
Format:
Adobe Portable Document Format
Description: