Response functions of a germanium-sodium iodide detector system

dc.contributor.authorAllen, BJen_AU
dc.contributor.authorBird, JRen_AU
dc.contributor.authorEngstrom, Sen_AU
dc.date.accessioned2007-11-22T04:30:36Zen_AU
dc.date.accessioned2010-04-30T04:40:29Zen_AU
dc.date.available2007-11-22T04:30:36Zen_AU
dc.date.available2010-04-30T04:40:29Zen_AU
dc.date.issued1967-04en_AU
dc.description.abstractA gamma ray spectrometer is described which uses a lithium drifted germanium diode or sodium iodide crystal as a central detector in conjunction with an annular segmented sodium iodide assembly. The system can operate as a total absorption, anticoincidence, or pair spectrometer and individual detectors may be used separately. Thus, the requirements of high resolution or high efficiency gamma ray spectroscopy can be met by suitable choice of mode of operation. The various modes of operation are compared and typical results given to illustrate their performance at a variety of gamma ray energies. A detailed analysis is given of the response of 30 cm3 Ge(Li) detector for gamma rays up to 17.6 MeV.en_AU
dc.identifier.citationAllen, B., Bird, J., & Engstrom, S. (1967). Response functions of a germanium-sodium iodide detector system (AAEC/TM380). Lucas Heights, N.S.W.: Research Establishment, Australian Atomic Energy Commission.en_AU
dc.identifier.govdoc802en_AU
dc.identifier.otherAAEC-TM-380en_AU
dc.identifier.placeofpublicationLucas Heights, New South Walesen_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/792en_AU
dc.language.isoen_auen_AU
dc.publisherAustralian Atomic Energy Commissionen_AU
dc.subjectGermaniumen_AU
dc.subjectSodium iodidesen_AU
dc.subjectCosmic photonsen_AU
dc.subjectSpectrometersen_AU
dc.titleResponse functions of a germanium-sodium iodide detector systemen_AU
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
AAEC-TM-380.pdf
Size:
507.99 KB
Format:
Adobe Portable Document Format
Description: