The Australian Commonwealth standard of measurement for absorbed radiation dose, Part 2 - absorbed dose standard for graphite irradiated by a cobalt-60 teletherapy unit
dc.contributor.author | Sherlock, SL | en_AU |
dc.date.accessioned | 2007-11-22T04:29:26Z | en_AU |
dc.date.accessioned | 2010-04-30T04:39:29Z | en_AU |
dc.date.available | 2007-11-22T04:29:26Z | en_AU |
dc.date.available | 2010-04-30T04:39:29Z | en_AU |
dc.date.issued | 1990-06 | en_AU |
dc.description.abstract | This report documents the absorbed dose standard for photon beams in the range from 1 to 25 MeV. Measurements of absorbed dose in graphite irradiated by a beam of cobalt-60 gamma rays from an Atomic Energy of Canada Limited (AECL) E1 Dorado 6 teletherapy unit are reported. The measurements were performed using a graphite calorimeter which is the primary standard for absorbed dose. The measurements are used to calibrate a working standard ion chamber in terms of absorbed dose in graphite. | en_AU |
dc.identifier.citation | Sherlock, S. L. (1990). The Australian Commonwealth standard of measurement for absorbed radiation dose, Part 2 - absorbed dose standard for graphite irradiated by a cobalt-60 teletherapy unit (ANSTO/E690). Lucas Heights, NSW: Australian Nuclear Science and Technology Organisation. | en_AU |
dc.identifier.govdoc | 746 | en_AU |
dc.identifier.isbn | 0642599041 | en_AU |
dc.identifier.issn | 10307745 | en_AU |
dc.identifier.other | ANSTO-E-690 | en_AU |
dc.identifier.placeofpublication | Lucas Heights, New South Wales | en_AU |
dc.identifier.uri | http://apo.ansto.gov.au/dspace/handle/10238/736 | en_AU |
dc.language.iso | en_au | en_AU |
dc.publisher | Australian Nuclear Science and Technology Organisation | en_AU |
dc.subject | Calorimetric dosemeters | en_AU |
dc.subject | Calibration standards | en_AU |
dc.subject | Australia | en_AU |
dc.subject | Ionization chambers | en_AU |
dc.subject | Radiation doses | en_AU |
dc.title | The Australian Commonwealth standard of measurement for absorbed radiation dose, Part 2 - absorbed dose standard for graphite irradiated by a cobalt-60 teletherapy unit | en_AU |