dc.description.abstract99mTc generators distributed by the A.A.E.C. are based on 99Mo extracted from irradiated uranium. The production of these generators and the methods employed in their testing are described. The quantity of 99mTc eluted from these generators varied with pH of the eluant with maxima at pH 0.5 - 0.9, 3, 4.7 - 6.7 and > 9.5. Terminal autoclaving reduced the yield by 10 - 20 per cent.en_AU
dc.identifier.citationBoyd, R. E., Hetherington, E. L. R., & Wood, N. R. (1971). Technetium-99m generators prepared from fission produced molybdenum-99- quality control and performance aspects (AAEC/E224). Lucas Heights, N.S.W.: Research Establishment, Australian Atomic Energy Commission.en_AU
dc.identifier.govdoc455en_AU
dc.identifier.isbn0642994188en_AU
dc.identifier.otherAAEC-E-224en_AU
dc.identifier.urihttp://apo.ansto.gov.au/dspace/handle/10238/412en_AU
dc.language.isoen_auen_AU
dc.publisherAustralian Atomic Energy Commissionen_AU
dc.subjectGamma spectrometersen_AU
dc.subjectIonization chambersen_AU
dc.subjectMolybdenum 99en_AU
dc.subjectTechnetium 99en_AU
dc.subjectIsotopesen_AU
dc.subjectFissionen_AU
dc.titleTechnetium-99m generators prepared from fission produced Molybdenum-99: quality control and performance aspectsen_AU
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
AAEC-E-224.pdf
Size:
909.54 KB
Format:
Adobe Portable Document Format
Description: