Please use this identifier to cite or link to this item: https://apo.ansto.gov.au/dspace/handle/10238/926
Title: Calibration and operation of the AAEC working standard of measurement for the activity of radionuclides, Part 2 - ion chamber analysis.
Authors: Wyllie, HA
Keywords: AAEC
Calibration standards
Cesium
Gamma radiation
Radioactivity
Issue Date: May-1986
Publisher: Australian Atomic Energy Commission
Citation: Wyllie, H. A. (1986). Calibration and operation of the AAEC working standard of measurement for the activity of radionuclides, Part 2 - ion chamber analysis (AAEC/E630). Lucas Heights, NSW: Australian Atomic Energy Commission.
Abstract: As a secondary method for the determination of radioactivity the ion chamber is both convenient and precise. The equations used in the calibration of the equipment and in the calculation of unknown activities are fully derived and it is shown that the experimental procedure outlined is the optimum for both precision and convenience.