Please use this identifier to cite or link to this item: https://apo.ansto.gov.au/dspace/handle/10238/237
Title: A direct reading spectrometer for the determination of beryllium in site liquid effluent.
Authors: Dale, LS
Keywords: Liquid wastes
Beryllium
Spectrometers
AAEC
Issue Date: Feb-1968
Publisher: Australian Atomic Energy Commission
Citation: Dale, L. S. (1968). A direct reading spectrometer for the determination of beryllium in site liquid effluent. (AAEC/TM438). Lucas Heights, NSW: Australian Atomic Energy Commission.
Abstract: This report describes a single channel direct reading spectrometer for the determination of beryllium in liquid effluent samples from delay tanks on the site of the Australian Atomic Energy Commission's Research Establishment. The rotating disc technique with pulsed arc excitation is used to determine beryllium down to 0.1 p.p.m. The limit of detection is 0.01 p.p.m. and the coefficient of variation at the 0.1 p.p.m. level is + 20 per cent for a 95 per cent confidence level. The beryllium intensity is depressed by the presence of other constituents in the effluent. Calibration is therefore carried out with standards prepared from a composite effluent solution which yields a similar depression of the beryllium intensity. The results are biased for different tanks and range from + 20 per cent to - 10 per cent. The nature of the depression of the beryllium intensity has been studied in relation to the interference effects of foreign anions and cations and the sensitivity of the method is shown to be affected by pH and by the grade of graphite discs used.
Gov't Doc #: 357
URI: http://apo.ansto.gov.au/dspace/handle/10238/237
Other Identifiers: AAEC-TM-438
Appears in Collections:Scientific and Technical Reports

Files in This Item:
File Description SizeFormat 
AAEC-TM-438.pdf520.04 kBAdobe PDFThumbnail
View/Open