Low-level gamma-spectrometry by using Compton suppression — testing the performance of two systems at the ANSTO low level radiochemistry laboratory

Loading...
Thumbnail Image
Date
2008-11-25
Journal Title
Journal ISSN
Volume Title
Publisher
South Pacific Radioactivity Association
Abstract
The successful use of gamma-spectrometry to measure very low level radioactivity relies on several factors, including the actual system and shielding, as well as environmental factors. A low-level laboratory needs to be in a geographic location with low back ground radiation, built with low background materials. The laboratory should ideally be built underground to minimize the interference from cosmic radiation. Compton suppressed systems have been designed to reduce background caused by Compton scattering. The background can be reduced by the use of Compton Suppression. This is achieved by introducing a second detection system. This second detection system, consisting of a large single crystal Nal annulus, detects the escaping energy from partial absorbed photons in the primary detector, the High purity Germanium detector. As set of standards tests to check the performance of the Compton suppression systems has b n devised by the Monaco Laboratory ( J-F Commanducci, 2003). These tests have been carried out on both ANSTO systems and show a considerable improvement in Compton to peak ratios in the testing and an overall decrease in background due to the shipping of incomplete photon absorption in the primary detector. The two systems allow small environmental samples to be analysed for isotopes such as Pb-210, Cs-137 and Be-7, without compromising the detection limit too much. J-F Commanducci, 2003: Construction du Laboratoire Souterrain et des Spectrometres a coincindences “faibre bruit” destines a la surveillance de L’Environrnent Marin. IAEA-MEL Document. PP 69.
Description
Keywords
Spectroscopy, Gamma spectrometers, Inhibition, Radioactivity, Shielding, ANSTO, Testing, GE semiconductor detectors
Citation
Heijnis, H., Zawadzki, A, W., & Fierro, D. (2008). Low-level gamma-spectrometry by using Compton suppression — testing the performance of two systems at the ANSTO low level radiochemistry laboratory. Paper presented to SPERA 08, South Pacific Environmental Radioactivity Association Conference, University of Canterbury, Christchurch, New Zealand, 23-27 November 2008. (pp. 52).