Browsing by Author "Dender, D"
Now showing 1 - 1 of 1
Results Per Page
Sort Options
- ItemSafety interlock and vent system to alleviate potentially dangerous ice blockage of top-loading cryostat sample sticks(Wiley-Blackwell, 2013-08-01) Pangelis, S; Olsen, SR; Scherschligt, J; Leao, JB; Pullen, SA; Dender, D; Hester, JR; Imperia, PA combined solution is presented for minimizing the safety hazards associated with closed cycle cryostats described by Swainson & Cranswick [J. Appl. Cryst. (2010), 43, 206-210]. The initial solution is to install a vent tube with one open end deep inside the sample space and a pressure relief valve at the top. This solution works for either a cryogen or a cryogen-free (closed cycle) system. The second approach, which can be combined with the first and is applicable to cryogen-free cryostats, involves electrically interlocking the closed cycle refrigerator compressor to the sample space, so that the system cannot be cooled in the presence of a leak path to air. © 2013, Wiley-Blackwell.