Studies of the characteristics of a silicon neutron sensor

No Thumbnail Available
Date
2009-08-18
Journal Title
Journal ISSN
Volume Title
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Abstract
Electrical characteristics and neutron dosimetry properties of silicon based p-i-n diodes are presented in support of the applications in the sensors for beam monitoring and medical physics. Both the current-voltage (I-V) and capacitance-voltage (C-V) characteristics of silicon planar p-i-n diode sensors with cylindrical geometry have been theoretically modeled and experimentally measured. The shifts of the forward and reverse diode characteristics of the sensors versus the neutron dose have been obtained. It is shown that the neutron irradiation caused shift of the forward voltage of the p-i-n diodes is proportional to the current at which it is measured in the case of the low level injection or to the square root of the current in the case of the high level injection. The C-V characteristics and the full depletion voltages of the diodes have been estimated and experimentally verified. It is shown that the sensitivity of planar cylindrical structures as neutron sensors can be optimized by the selection of the device geometry and the current at which the measurement is performed. © 2009, Institute of Electrical and Electronics Engineers (IEEE)
Description
Keywords
Silicon, Neutron dosimetry, Electric potential, Irradiation, Radiation detectors, Electron density
Citation
Anokhin, I., Zinets, O., Rosenfeld, R., Lerch, M., Yudelev, M., Perevertaylo, V., Reinhard, M., & Petasecca, M. (2009). Studies of the characterisation of a silicon neutron sensor. IEEE Transactions on Nuclear Science, 56(4), 2290-2293. doi:10.1109/TNS.2009.2024150
Collections