ANSTO Publications Online >
Journal Publications >
Journal Articles >
Please use this identifier to cite or link to this item:
http://apo.ansto.gov.au/dspace/handle/10238/1076
|
| Title: | Evidence of a water layer in solid-contact polymeric ion sensors. |
| Authors: | De Marco, R Veder, JP Clarke, G Nelson, A Prince, KE Pretsch, E Bakker, E |
| Keywords: | Transducers Ion-Selective Electrodes Neutral Currents Electrodes Membranes Polymers Sensitivity |
| Issue Date: | Jan-2008 |
| Publisher: | Royal Society of Chemistry |
| Citation: | De Marco, R., Veder, J. P., Clarke, G., Nelson, A., Prince, K., Pretsch, E., et al. (2008). Evidence of a water layer in solid-contact polymeric ion sensors. Physical Chemistry Chemical Physics, 10(1), 73-76. |
| Abstract: | This paper presents the very first direct structural evidence for the formation of a 100 +/- 10 angstrom water layer in coated-wire polymeric-membrane ion-selective electrodes (ISEs). © 2008, Royal Society of Chemistry |
| URI: | http://dx.doi.org/10.1039/b714248j http://apo.ansto.gov.au/dspace/handle/10238/1076 |
| ISSN: | 1463-9076 |
| Appears in Collections: | Journal Articles
|
Files in This Item:
There are no files associated with this item.
|